Vulnerability · Published Jun 2, 2026 · Assigned by VulnCheck
CVE-2025-15653Dräger Zeus IE Anesthesia Workstation USB Interface Privilege Escalation
Medium 6.8
In plain English
A medium-severity security flaw in Zeus IE from Dräger. There are no confirmed reports of exploitation so far. The chance it is exploited in the next 30 days is estimated at 0.2% (higher than 6% of all known vulnerabilities).
Official description
Dräger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dräger Service Connect.
Affected products (as reported)
| Vendor | Product | Versions |
|---|---|---|
| Dräger | Zeus IE | 0 – before 1.0.5 |
| Dräger | Zeus RS C500 | 0 – before 1.0.9 |
Weakness: CWE-668 · Vector: CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H