Vulnerability · Published Nov 14, 2024 · Assigned by Baxter
CVE-2024-48974Life2000 Ventilator does not perform proper file integrity checks when adopting firmware updates
Critical 9.3
In plain English
A critical-severity security flaw in Life2000 Ventilation System from Baxter. There are no confirmed reports of exploitation so far. The chance it is exploited in the next 30 days is estimated at 0.2% (higher than 5% of all known vulnerabilities).
Official description
The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.
Affected products (as reported)
| Vendor | Product | Versions |
|---|---|---|
| Baxter | Life2000 Ventilation System | 06.08.00.00 and prior |
Weakness: CWE-494 · Vector: CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H