Vulnerability · Published Nov 14, 2024 · Assigned by Baxter

CVE-2024-48966Life2000 service tools for test and calibration do not support user authentication

Critical 10.0

In plain English

A critical-severity missing authentication in Life2000 Ventilation System from Baxter. There are no confirmed reports of exploitation so far. The chance it is exploited in the next 30 days is estimated at 0.7% (higher than 53% of all known vulnerabilities).

Severity (CVSS 3.1)10.0Critical
Exploit odds (30 days)0.7%top 46.8% of all CVEs
Exploited in the wildNo reportsCISA SSVC: none
Attack toolingNoneno public tooling known

Official description

The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

Affected products (as reported)

VendorProductVersions
Baxter Life2000 Ventilation System 06.08.00.00 and prior

Weakness: CWE-306 · Vector: CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H