Vulnerability · Published Mar 1, 2023 · Assigned by Medtronic
CVE-2023-25931Medtronic Micro Clinician & InterStim X Clinician App Password Reset Issue
Medium 6.4
In plain English
A medium-severity security flaw in InsterStim Applications from Medtronic. There are no confirmed reports of exploitation so far. The chance it is exploited in the next 30 days is estimated at 0.3% (higher than 19% of all known vulnerabilities).
Official description
Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. Changes still cannot be made outside of the established therapy parameters of the programmer. For unauthorized access to occur, an individual would need physical access to the Smart Programmer.
Affected products (as reported)
| Vendor | Product | Versions |
|---|---|---|
| Medtronic | InsterStim Applications | Micro Clinician – before A51200InterStim X Clinician – before A51300 |
Weakness: CWE-620 · Vector: CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:H