Vulnerability · Published Mar 13, 2023 · Assigned by B.Braun
CVE-2023-0888Authenticated eval injection in B. Braun Space Battery pack SP with Wi-Fi
Medium 4.9
In plain English
A medium-severity security flaw in Space Battery Pack SP with Wi-Fi from B. Braun . There are no confirmed reports of exploitation so far. The chance it is exploited in the next 30 days is estimated at 0.6% (higher than 49% of all known vulnerabilities).
Official description
An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks
Affected products (as reported)
| Vendor | Product | Versions |
|---|---|---|
| B. Braun | Space Battery Pack SP with Wi-Fi | 0 – through 053L0000920 – through 054U000092 |
Weakness: CWE-95 · Vector: CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:N/I:H/A:N