Vulnerability · Published Sep 13, 2022 · Assigned by icscert
CVE-2022-38100Contec Health CMS8000
High 7.5
In plain English
A high-severity uncontrolled resource consumption in CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor from Contec Health. There are no confirmed reports of exploitation so far. The chance it is exploited in the next 30 days is estimated at 1.0% (higher than 62% of all known vulnerabilities).
Official description
The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network.
Affected products (as reported)
| Vendor | Product | Versions |
|---|---|---|
| Contec Health | CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor | All |
Weakness: CWE-400 · Vector: CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
References
- https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01 x_refsource_MISC