Vulnerability · Published Jun 24, 2022 · Assigned by icscert
CVE-2022-15243.2.5 CLEARTEXT TRANSMISSION OF SENSITIVE INFORMATION CWE-319
High 7.4
In plain English
A high-severity security flaw in iSeq 100 Instrument from Illumina. There are no confirmed reports of exploitation so far. The chance it is exploited in the next 30 days is estimated at 0.3% (higher than 26% of all known vulnerabilities).
Severity (CVSS 3.1)7.4High
Exploit odds (30 days)0.3%top 74.4% of all CVEs
Exploited in the wildNo reportsCISA SSVC: none
Attack toolingNoneno public tooling known
Official description
LRM version 2.4 and lower does not implement TLS encryption. A malicious actor can MITM attack sensitive data in-transit, including credentials.
Affected products (as reported)
| Vendor | Product | Versions |
|---|---|---|
| Illumina | NextSeq 550Dx | LRM Versions 1.3 to 3.1 |
| Illumina | MiSeq Dx | LRM Versions 1.3 to 3.1 |
| Illumina | NextSeq 500 Instrument | LRM Versions 1.3 to 3.1 |
| Illumina | NextSeq 550 Instrument | LRM Versions 1.3 to 3.1 |
| Illumina | MiSeq Instrument | LRM Versions 1.3 to 3.1 |
| Illumina | iSeq 100 Instrument | LRM Versions 1.3 to 3.1 |
| Illumina | MiniSeq Instrument | LRM Versions 1.3 to 3.1 |
Weakness: CWE-319 · Vector: CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:N
References
- https://www.cisa.gov/uscert/ics/advisories/icsa-22-153-02 x_refsource_MISC