CVE-2025-41690Endress+Hauser: Proline 10 Maintenance credentials may be exposed under certain conditions

High 7.4

In plain English

A high-severity sensitive data in logs in Promag 10 with HART from Endress+Hauser. There are no confirmed reports of exploitation so far. The chance it is exploited in the next 30 days is estimated at 0.2% (higher than 15% of all known vulnerabilities).

Severity (CVSS 3.1)7.4High
Exploit odds (30 days)0.2%top 85.4% of all CVEs
Exploited in the wildNo reportsCISA SSVC: none
Attack toolingNoneno public tooling known

Official description

A low-privileged attacker in bluetooth range may be able to access the password of a higher-privilege user (Maintenance) by viewing the device’s event log. This vulnerability could allow the Operator to authenticate as the Maintenance user, thereby gaining unauthorized access to sensitive configuration settings and the ability to modify device parameters.

Affected products (as reported)

VendorProductVersions
Endress+Hauser Promag 10 with HART 0 – before 01.00.06
Endress+Hauser Promag 10 with IO-Link 0 – before 01.00.02
Endress+Hauser Promag 10 with Modbus 0 – before 01.00.06
Endress+Hauser Promass 10 with HART 0 – before 01.00.06
Endress+Hauser Promass 10 with IO-Link 0 – before 01.00.02
Endress+Hauser Promass 10 with Modbus 0 – before 01.00.06

Weakness: CWE-532 · Vector: CVSS:3.1/AV:A/AC:L/PR:L/UI:R/S:U/C:H/I:H/A:H