Vulnerability · Published Dec 17, 2025 · Assigned by Radiometer
CVE-2025-14096Credential Disclosure vulnerability in Radiometer Products
High 8.4
In plain English
A high-severity hard-coded credentials in ABL800 BASIC and ABL800 FLEX analyzers from Radiometer Medical Aps. There are no confirmed reports of exploitation so far. The chance it is exploited in the next 30 days is estimated at 0.1% (higher than 3% of all known vulnerabilities).
Official description
A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system. Other related CVE's are CVE-2025-14095 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Attacker requires physical access to the analyzer. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.
Affected products (as reported)
| Vendor | Product | Versions |
|---|---|---|
| Radiometer Medical Aps | ABL90 FLEX and ABL90 FLEX PLUS Analyzers | Windows 7 Operating systemWindows XP Operating system |
| Radiometer Medical Aps | AQT90 FLEX Analyzers | Windows 7 Operating systemWindows XP Operating system |
| Radiometer Medical Aps | ABL800 BASIC and ABL800 FLEX analyzers | Windows 7 Operating systemWindows XP Operating system |
Weakness: CWE-798, CWE-250 · Vector: CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H