In plain English

A medium-severity improper access control in ABL800 BASIC and ABL800 FLEX Analyzers from Radiometer Medical Aps. There are no confirmed reports of exploitation so far. The chance it is exploited in the next 30 days is estimated at 0.2% (higher than 6% of all known vulnerabilities).

Severity (CVSS 3.1)6.8Medium
Exploit odds (30 days)0.2%top 94.0% of all CVEs
Exploited in the wildNo reportsCISA SSVC: none
Attack toolingNoneno public tooling known

Official description

A "Privilege boundary violation" vulnerability is identified affecting multiple Radiometer Products. Exploitation of this vulnerability gives a user with physical access to the analyzer, the possibility to gain unauthorized access to functionalities outside the restricted environment. The vulnerability is due to weakness in the design of access control implementation in application software.  Other related CVE's are CVE-2025-14096 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required configuration for Exposure: Physical access to the analyzer is needed. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided working proof-of-concept. Radiometer is not aware of any publicly available exploit at the time of publication.                                                                                                                                                                                        Note: CVSS score 6.8 when underlying OS is Windows 7 or Windows XP Operating systems and CVSS score 5.7 when underlying OS is Windows 8 or Windows 10 operating systems.

Affected products (as reported)

VendorProductVersions
Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers All application software versions with Windows 7, Windows XP as underlying OSApplication software versions < 3.5MR11 with Windows 10 as underlying OS
Radiometer Medical Aps AQT90 FLEX Analyzers All Application software versions <= 8.13 MR2
Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers Application software versions < 6.20 MR2 with Windows 7, Windows XP as underlying OSApplication software versions < 6.20 MR2 with Windows 10 as underlying OS
Radiometer Medical Aps ABL9 Analyzers Application software versions < 1.5.0

Weakness: CWE-284, CWE-693 · Vector: CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H