Vulnerability · Published Feb 25, 2025 · Assigned by icscert
CVE-2025-1204
High 7.7
In plain English
A high-severity security flaw in CMS8000 Patient Monitor from Contec Health. Proof-of-concept exploit code is public. The chance it is exploited in the next 30 days is estimated at 0.5% (higher than 42% of all known vulnerabilities).
Official description
The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the 'C' button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device.
Affected products (as reported)
| Vendor | Product | Versions |
|---|---|---|
| Contec Health | CMS8000 Patient Monitor | 0 |
Weakness: CWE-912 · Vector: CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N