Vulnerability · Published Jan 30, 2025 · Assigned by icscert

CVE-2025-0626Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor

High 7.5

In plain English

A high-severity security flaw in CMS8000 Patient Monitor from Contec Health. There are no confirmed reports of exploitation so far. The chance it is exploited in the next 30 days is estimated at 1.1% (higher than 67% of all known vulnerabilities).

Severity (CVSS 3.1)7.5High
Exploit odds (30 days)1.1%top 33.2% of all CVEs
Exploited in the wildNo reportsCISA SSVC: none
Attack toolingNoneno public tooling known

Official description

The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.

Affected products (as reported)

VendorProductVersions
Contec Health CMS8000 Patient Monitor All versions

Weakness: CWE-912 · Vector: CVSS:3.1/AV:N/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:H