CVE-2024-48967Life2000 ventilator and Service PC lack sufficient audit logging capabilities
Critical 10.0
In plain English
A critical-severity security flaw in Life2000 Ventilation System from Baxter. There are no confirmed reports of exploitation so far. The chance it is exploited in the next 30 days is estimated at 0.6% (higher than 50% of all known vulnerabilities).
Official description
The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance.
Affected products (as reported)
| Vendor | Product | Versions |
|---|---|---|
| Baxter | Life2000 Ventilation System | 06.08.00.00 and prior |
Weakness: CWE-778 · Vector: CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H