Vulnerability · Published Dec 5, 2022 · Assigned by BD
CVE-2022-43557BD BodyGuard™ Pumps – RS-232 Interface Vulnerability
Medium 5.3
In plain English
A medium-severity security flaw in BodyGuard™ Pump from Becton, Dickinson and Company (BD). There are no confirmed reports of exploitation so far. The chance it is exploited in the next 30 days is estimated at 0.2% (higher than 13% of all known vulnerabilities).
Official description
The BD BodyGuard™ infusion pumps specified allow for access through the RS-232 (serial) port interface. If exploited, threat actors with physical access, specialized equipment and knowledge may be able to configure or disable the pump. No electronic protected health information (ePHI), protected health information (PHI) or personally identifiable information (PII) is stored in the pump.
Affected products (as reported)
| Vendor | Product | Versions |
|---|---|---|
| Becton, Dickinson and Company (BD) | BodyGuard™ Pump | BD BodyGuard™ CME BodyGuard™ 323 (2nd Edition)CME BodyGuard™ 323 Color Vision (2nd Edition)CME BodyGuard™ 323 Color Vision (3rd Edition)CME BodyGuard™ Twins (2nd Edition) |
Weakness: CWE-1299 · Vector: CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:H