Vulnerability · Published Aug 2, 2021 · Assigned by mitre

CVE-2021-37164

Critical 9.8

In plain English

A critical-severity security flaw in the affected software. There are no confirmed reports of exploitation so far. The chance it is exploited in the next 30 days is estimated at 3.4% (higher than 89% of all known vulnerabilities).

Severity (CVSS NVD)9.8Critical
Exploit odds (30 days)3.4%top 11.1% of all CVEs
Exploited in the wildNo reportsnot in CISA KEV
Attack toolingNoneno public tooling known

Official description

A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. In the tcpTxThread function, the received data is copied to a stack buffer. An off-by-3 condition can occur, resulting in a stack-based buffer overflow.

Affected products (as reported)

VendorProductVersions
n/a n/a n/a

References