Vulnerability · Published Aug 2, 2021 · Assigned by mitre
CVE-2021-37163
Critical 9.8
In plain English
A critical-severity security flaw in the affected software. There are no confirmed reports of exploitation so far. The chance it is exploited in the next 30 days is estimated at 1.4% (higher than 73% of all known vulnerabilities).
Severity (CVSS NVD)9.8Critical
Exploit odds (30 days)1.4%top 26.7% of all CVEs
Exploited in the wildNo reportsnot in CISA KEV
Attack toolingNoneno public tooling known
Official description
An insecure permissions issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus operated by released versions of software before Nexus Software 7.2.5.7. The device has two user accounts with passwords that are hardcoded.
Affected products (as reported)
References
- https://www.swisslog-healthcare.com x_refsource_MISC
- https://www.armis.com/PwnedPiper x_refsource_MISC
- https://www.swisslog-healthcare.com/en-us/customer-care/security-information/cve-disclosures#:~:text=CVE%20Disclosures%20%20%20%20Vulnerability%20Name%20%2C%20%20CVE-2021-37164%20%204%20more%20rows%20 x_refsource_MISC
- https://www.swisslog-healthcare.com/-/media/swisslog-healthcare/documents/customer-service/armis-documents/cve-2021-37163-bulletin---default-credentials-for-the-telnet-server.pdf?rev=da64c389a475494985b9fd2c2c508542&hash=466A7109AF08EBFF3756B2C25968ED5E x_refsource_MISC