Vulnerability · Published Jan 21, 2022 · Assigned by icscert

CVE-2021-33846Fresenius Kabi Agilia Connect Infusion System use of a broken or risky cryptographic algorithm

Medium 5.9

In plain English

A medium-severity broken cryptography in Vigilant Software Suite (Mastermed Dashboard) from Fresenius Kabi. There are no confirmed reports of exploitation so far. The chance it is exploited in the next 30 days is estimated at 0.3% (higher than 24% of all known vulnerabilities).

Severity (CVSS 3.1)5.9Medium
Exploit odds (30 days)0.3%top 75.6% of all CVEs
Exploited in the wildNo reportsCISA SSVC: none
Attack toolingNoneno public tooling known

Official description

Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 issues authentication tokens to authenticated users that are signed with a symmetric encryption key. An attacker in possession of the key can issue valid JWTs and impersonate arbitrary users.

Affected products (as reported)

VendorProductVersions
Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) unspecified – before 2.0.1.3

Weakness: CWE-327 · Vector: CVSS:3.1/AV:N/AC:H/PR:H/UI:N/S:U/C:H/I:H/A:N

References