Vulnerability · Published Mar 15, 2021 · Assigned by icscert
CVE-2020-27278
Medium 5.2
In plain English
A medium-severity hard-coded credentials in Hamilton Medical AG, T1-Ventillator. There are no confirmed reports of exploitation so far. The chance it is exploited in the next 30 days is estimated at 0.3% (higher than 21% of all known vulnerabilities).
Severity (CVSS NVD)5.2Medium
Exploit odds (30 days)0.3%top 79.2% of all CVEs
Exploited in the wildNo reportsnot in CISA KEV
Attack toolingNoneno public tooling known
Official description
In Hamilton Medical AG,T1-Ventillator versions 2.2.3 and prior, hard-coded credentials in the ventilator allow attackers with physical access to obtain admin privileges for the device’s configuration interface.
Affected products (as reported)
| Vendor | Product | Versions |
|---|---|---|
| n/a | Hamilton Medical AG, T1-Ventillator | Versions 2.2.3 and prior |
Weakness: CWE-798
References
- https://us-cert.cisa.gov/ics/advisories/icsma-21-047-01 x_refsource_MISC