Vulnerability · Published Mar 15, 2021 · Assigned by icscert

CVE-2020-27278

Medium 5.2

In plain English

A medium-severity hard-coded credentials in Hamilton Medical AG, T1-Ventillator. There are no confirmed reports of exploitation so far. The chance it is exploited in the next 30 days is estimated at 0.3% (higher than 21% of all known vulnerabilities).

Severity (CVSS NVD)5.2Medium
Exploit odds (30 days)0.3%top 79.2% of all CVEs
Exploited in the wildNo reportsnot in CISA KEV
Attack toolingNoneno public tooling known

Official description

In Hamilton Medical AG,T1-Ventillator versions 2.2.3 and prior, hard-coded credentials in the ventilator allow attackers with physical access to obtain admin privileges for the device’s configuration interface.

Affected products (as reported)

VendorProductVersions
n/a Hamilton Medical AG, T1-Ventillator Versions 2.2.3 and prior

Weakness: CWE-798

References