Vulnerability · Published Aug 26, 2020 · Assigned by mitre

CVE-2020-15486

Medium 6.5

In plain English

A medium-severity security flaw in the affected software. There are no confirmed reports of exploitation so far. The chance it is exploited in the next 30 days is estimated at 0.5% (higher than 44% of all known vulnerabilities).

Severity (CVSS NVD)6.5Medium
Exploit odds (30 days)0.5%top 56.4% of all CVEs
Exploited in the wildNo reportsnot in CISA KEV
Attack toolingNoneno public tooling known

Official description

An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of the device and can sniff the data being broadcasted while a measurement is being done. Also, saved data can also be extracted over a Bluetooth connection. In addition, an attacker can launch a man-in-the-middle attack against data integrity.

Affected products (as reported)

VendorProductVersions
n/a n/a n/a

References