Vulnerability · Published Jun 29, 2020 · Assigned by icscert
CVE-2020-12048
High 7.5
In plain English
A high-severity security flaw in Baxter Phoenix Hemodialysis Delivery System. There are no confirmed reports of exploitation so far. The chance it is exploited in the next 30 days is estimated at 0.5% (higher than 39% of all known vulnerabilities).
Official description
Phoenix Hemodialysis Delivery System SW 3.36 and 3.40, The Phoenix Hemodialysis device does not support data-in-transit encryption (e.g., TLS/SSL) when transmitting treatment and prescription data on the network between the Phoenix system and the Exalis dialysis data management tool. An attacker with access to the network could observe sensitive treatment and prescription data sent between the Phoenix system and the Exalis tool.
Affected products (as reported)
| Vendor | Product | Versions |
|---|---|---|
| n/a | Baxter Phoenix Hemodialysis Delivery System | Phoenix Hemodialysis Delivery System SW 3.36 and 3.40 |
Weakness: CWE-319
References
- https://www.us-cert.gov/ics/advisories/icsma-20-170-03 x_refsource_MISC