Vulnerability · Published Jun 29, 2020 · Assigned by icscert

CVE-2020-12041

Critical 9.4

In plain English

A critical-severity incorrect permission assignment in Baxter Sigma Spectrum Infusion Pumps. There are no confirmed reports of exploitation so far. The chance it is exploited in the next 30 days is estimated at 1.4% (higher than 73% of all known vulnerabilities).

Severity (CVSS NVD)9.4Critical
Exploit odds (30 days)1.4%top 27.3% of all CVEs
Exploited in the wildNo reportsnot in CISA KEV
Attack toolingNoneno public tooling known

Official description

The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to sensitive data stored on the WBM that permits temporary configuration changes to network settings of the WBM, and allows the WBM to be rebooted. Temporary configuration changes to network settings are removed upon reboot.

Affected products (as reported)

VendorProductVersions
n/a Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum v6.x model 35700BAX, Baxter Spectrum v8.x model 35700BAX2,Sigma Spectrum v6.x with Wireless Battery Module v9,11,13,14,15,16,v20D29,v20D30,v20D31,v22D24, Baxter Spectrum v8.x with Wireless Battery Module v17,v20D29,v20D30,v20D31,v22D24,Baxter Spectrum Wireless Battery Module v17,v20D29,v20D30,v20D31,v22D24,Baxter Spectrum LVP v8.x w/Wireless Battery Module v17,v20D29,v20D30,v20D31,v22D24

Weakness: CWE-732