Vulnerability · Published Jun 29, 2020 · Assigned by icscert
CVE-2020-12037
High 7.5
In plain English
A high-severity security flaw in Baxter PrismaFlex and PrisMax. There are no confirmed reports of exploitation so far. The chance it is exploited in the next 30 days is estimated at 0.5% (higher than 42% of all known vulnerabilities).
Severity (CVSS NVD)7.5High
Exploit odds (30 days)0.5%top 58.2% of all CVEs
Exploited in the wildNo reportsnot in CISA KEV
Attack toolingNoneno public tooling known
Official description
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.
Affected products (as reported)
| Vendor | Product | Versions |
|---|---|---|
| n/a | Baxter PrismaFlex and PrisMax | PrismaFlex all versions, PrisMax all versions prior to 3.x |
Weakness: CWE-259
References
- https://www.us-cert.gov/ics/advisories/icsma-20-170-01 x_refsource_MISC