A medium-severity improper authentication in Baxter PrismaFlex and PrisMax. There are no confirmed reports of exploitation so far. The chance it is exploited in the next 30 days is estimated at 0.2% (higher than 16% of all known vulnerabilities).

Severity (CVSS NVD)4.9Medium
Exploit odds (30 days)0.2%top 83.9% of all CVEs
Exploited in the wildNo reportsnot in CISA KEV
Attack toolingNoneno public tooling known

Official description

Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The PrismaFlex device contains a hard-coded service password that provides access to biomedical information, device settings, calibration settings, and network configuration. This could allow an attacker to modify device settings and calibration.

Affected products (as reported)

VendorProductVersions
n/a Baxter PrismaFlex and PrisMax PrismaFlex all versions, PrisMax all versions prior to 3.x

Weakness: CWE-287

References