Vulnerability · Published Jun 29, 2020 · Assigned by icscert
CVE-2020-12024
Medium 6.1
In plain English
A medium-severity improper access control in Baxter ExactaMix EM 2400 & EM 1200. There are no confirmed reports of exploitation so far. The chance it is exploited in the next 30 days is estimated at 0.4% (higher than 30% of all known vulnerabilities).
Official description
Baxter ExactaMix EM 2400 versions 1.10, 1.11, 1.13, 1.14 and ExactaMix EM1200 Versions 1.1, 1.2, 1.4 and 1.5 does not restrict access to the USB interface from an unauthorized user with physical access. Successful exploitation of this vulnerability may allow an attacker with physical access to the system the ability to load an unauthorized payload or unauthorized access to the hard drive by booting a live USB OS. This could impact confidentiality and integrity of the system and risk exposure of sensitive information including PHI.
Affected products (as reported)
| Vendor | Product | Versions |
|---|---|---|
| n/a | Baxter ExactaMix EM 2400 & EM 1200 | ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5 |
Weakness: CWE-284
References
- https://www.us-cert.gov/ics/advisories/icsma-20-170-01 x_refsource_MISC