Vulnerability · Published Jun 29, 2020 · Assigned by icscert
CVE-2020-12012
Medium 6.1
In plain English
A medium-severity security flaw in Baxter ExactaMix EM 2400 & EM 1200. There are no confirmed reports of exploitation so far. The chance it is exploited in the next 30 days is estimated at 0.4% (higher than 30% of all known vulnerabilities).
Official description
Baxter ExactaMix EM 2400 & EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13, and ExactaMix EM1200 Versions 1.1, 1.2, and 1.4 have hard-coded administrative account credentials for the ExactaMix application. Successful exploitation of this vulnerability may allow an attacker with physical access to gain unauthorized access to view/update system configuration or data. This could impact confidentiality and integrity of the system and risk exposure of sensitive information including PHI.
Affected products (as reported)
| Vendor | Product | Versions |
|---|---|---|
| n/a | Baxter ExactaMix EM 2400 & EM 1200 | ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5 |
Weakness: CWE-259
References
- https://www.us-cert.gov/ics/advisories/icsma-20-170-01 x_refsource_MISC