Vulnerability · Published Dec 1, 2021 · Assigned by icscert
CVE-2020-10627
High 7.3
In plain English
A high-severity improper access control in Omnipod Insulin Management System from Insulet. There are no confirmed reports of exploitation so far. The chance it is exploited in the next 30 days is estimated at 0.5% (higher than 42% of all known vulnerabilities).
Official description
Insulet Omnipod Insulin Management System insulin pump product ID 19191 and 40160 is designed to communicate using a wireless RF with an Insulet manufactured Personal Diabetes Manager device. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with access to one of the affected insulin pump models may be able to modify and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.
Affected products (as reported)
| Vendor | Product | Versions |
|---|---|---|
| Insulet | Omnipod Insulin Management System | 1919140160ZXP425ZXR425 |
Weakness: CWE-284 · Vector: CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:L/I:H/A:L
References
- https://us-cert.cisa.gov/ics/advisories/icsma-20-079-01 x_refsource_MISC
- https://www.myomnipod.com/security-bulletins x_refsource_MISC